A significant advantage is that all filters are replaceable without dismounting the probe without using any tools and in the shortest possible time. Cleaning and exchanging of unheated sample tubes or preliminary top-filters can be affected by extracting the filter from the probe.
The probe temperature is controlled by a microprocessor based PID-controller (optional with Modbus/RS485 communication). Alarm or fault contacts can be programmed and the temperature can be changed easily. The standard sensor is PT100, whereas a thermo-couple is standard for the high temperature version.
The following features are offered for all probes:
• Test gas can be injected directly into the probe according to EN14181 (regulation for calibration of emission monitoring systems) that enables calibration gas feeding via the filter element of the gas sample probe.
• Test gas can be injected into the probe through a check valve directly to the sample outlet so that no calibration gas is lost to the stack.
• An isolation valve with pneumatic control shuts off the sample outlet from the internal filter area in case of blow-back.
• Cleaning of filter and the sample tube through a high-flow inlet ports so less maintenance is necessary in high dust load applications. This inlet can be controlled by pneumatic or electric valves, and also in combination with a volume chamber for high pressure flow.